ETAC SIR13 / SIR13 mini / SIR13 SLIMInsulation degradation measurement system
  • Summary
  • Parameter
  • Application
  • Data
  • Related
  • Brand
  • Video
     SIR13 / SIR13 mini / SIR13 SLIM Features:

    ・The system is designed for 3-way us, namely Electro-chemical Migration detection, insulation resistance measurement and insulation characteristics evaluation.

    ・Ammeter is adopted to all CH measures insulation resistance not only continuously but also high-speed (20mse per CH). This system detects Electro-chemical Migration, which started and terminated instantly, accurately.

    ・Two types of 250V standard measurement board with 16CH are available. The one is 1ch/1power supply and another is 8ch/1 power supply. 500V and 1KV measurement board is 1ch/1 power supply. Depends on application, useful board is selected.

    ・SIR13 accepts to adopt different applied voltage measurement in one system. When Electro-chemical Migration was happened, previous function acquires both data before and after the phenomenon.

    ・By utilizing high power unit, insulation evaluation with applied voltage maximum. 10KV is possible.

    ・Pulse board, which can set any middle voltage between plus 100V and minas 100V and apply pulse voltage ware repeatedly, is added.

    ・± pulse applied wave can be set by 20 and any ware. It is possible to apply various pulse wave.


     SIR13 / SIR13 mini / SIR13 SLIM Specification:

    Applied   voltage

    100VPulse・250V・500V・1kV

    High voltage   unit

    1.5kV・3kV・6kV・10kV

    Measurement   board

    Max.8 board (in case of 16ch   board,Max128ch)

    Maximum   current measurement         range and resolution

    1)   30mA(1μA resolution)
     
      2) 3.2mA(10nA resolution)
     
      3) 320μA(1nA resolution)
      4) 3.2μA(10pA resolution)
     
      5) 32nA(100fA resolution)
     
      6) 320pA(1fA resolution)